Abstract

Thin-film-based electrodes used to interact with nervous tissue often fail quickly if used for electrical stimulation, impairing their translation into long-term clinical applications. We initiated investigations about the mechanical load on thin-film electrodes caused by the fact of electrical stimulation. Platinum electrodes of Ø 300μm on a polyimide carrier were subjected to approximately 50 000 asymmetrical, biphasic stimulation pulses in vitro. The electrode's surface was investigated optically by means of white-light interferometry. The structural expansion for the metallic surface subjected to stimulation was measured to reach roughly 30%. The study points towards a failure mechanism of thin-films being of mechanical nature, inherent to the unavoidable electrochemical processes involved (change in lattice constants) during electrical stimulation at the electrode's surface. Based on further scientific facts, we set 3 hypotheses for the exact mechanisms involved in the failure of thin-films used for electrical stimulation, opening a new door for research and improvement of novel neuroprosthetic devices.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.