Abstract

Wavefront analysis using a scanning slit refractometer (the OPD-Scan™) was evaluated in both artificial and human eyes, in comparison with the Hartmann-Shack wavefront sensor (H-S) currently often used. Comparison of different methods and configurations carried out in the artificial eyes yielded basically the same results in moderate refractive error cases. There was a closer match between the RMS (root mean square) wavefront error obtained by OPD-Scan and H-S in the 6 mm pupils than in the 4 mm pupils of the same normal human eyes. Although OPD-Scan employs a different approach in determining aberration, the aberration value is similar to that of H-S, especially in large pupils.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.