Abstract

This study aimed to compare ocular wavefront aberrations for pupil diameters of 4 mm and 6 mm, and contrast sensitivity, in eyes with AcrySof IQ and AcrySof Natural intraocular lenses (IOLs). Sixty eyes of 60 patients were enrolled in this prospective randomized study. After phacoemulsification the eyes received either AcrySof IQ SN60WF or AcrySof Natural SN60AT IOLs. One month after surgery, all patients underwent complete ophthalmological examination including corneal topography, wavefront analysis for pupil diameters of 4 mm and 6 mm, and contrast sensitivity measurements with the CSV 1000E instrument under photopic and mesopic conditions with and without glare. There was no statistically significant difference between groups in age, sex or other preoperative ocular characteristics (p > 0.05). Patients with AcrySof IQ IOLs had higher contrast sensitivity at 6 c.p.d. under photopic conditions, at 6 c.p.d. and 18 c.p.d. under mesopic conditions, and at 6 c.p.d., 12 c.p.d. and 18 c.p.d. under mesopic conditions with glare (p < 0.05). Corneal spherical aberration was 0.273 ± 0.074 μm in the AcrySof Natural group and 0.294 ± 0.086 μm in the AcrySof IQ group (p = 0489). Ocular spherical aberration was 0.362 ± 0.141 μm and 0.069 ± 0.043 μm (p < 0.001) for 6-mm diameter pupils and 0.143 ± 0.091 μm and 0.017 ± 0.016 μm (p < 0.001) for 4-mm diameter pupils, with AcrySof Natural and AcrySof IQ IOLs, respectively. There were no significant differences in other higher-order aberrations between the groups (p > 0.05). Aspherical AcrySof IQ IOLs significantly reduced spherical aberration for pupil diameters of both 4 mm and 6 mm and also improved contrast sensitivity more than spherical AcrySof Natural IOLs, especially in mesopic conditions.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call