Abstract

An adiabatic laser calorimeter for measuring weak absorption in optical thin films is presented. It consists of three main parts which are used to control and measure temperature, and for calibration. The controlling temperature part is a set of automatic temperature controls mainly composed of a PID adjuster. The temperature difference between specimen and environment is always less than +/-0.006 degrees C during the measurement process. There is almost no energy exchange between specimen and environment. The specimen temperature increment is measured by a high precision thermistor set. Its measuring temperature sensitivity is 0.003 degrees C and the accuracy is +/-0.004 degrees C. The absolute electrical calibration part is used to obtain absorptance of a specimen immediately without needing to know the mass and specific heat of the specimen. Absorptance as small as 1 x10(-5) can be detected when a 1.5-W argon-ion laser is used.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.