Abstract

This paper determines the optimum range of total ion current for a cesium liquid metal ion source for use as a focused ion beam (FIB). This range is determined from a figure of merit calculated from measurements of the angular intensity and energy spread of emitted ions. Judging from both the figure of merit and the tail of the energy distribution curves, the total emission current should be set near 1 μA for a cesium FIB with a high current density. Assuming that both Cs- and Ga-FIBs have the same diameter, the relative current density of a Cs-FIB is expected to be approximately 80% that of a Ga-FIB.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.