Abstract

The relationship between experimentally induced intraocular lens (IOL) haptic deformation and resulting elastic haptic counter-resisting forces measured by electronic dynamometry was examined for 34 different IOL haptics of varying material composition and geometrical designs. Poly(methyl methacrylate) (PMMA) and polypropylene loops of similar geometry did not fundamentally differ from one another, although lenses of differing geometry behaved differently. Unlike PMMA and polypropylene loops, soft haptics of poly-HEMA and silicone rubber demonstrated a larger elastic resistance force to the same degree of deformation. This was based upon design characteristics of the lenses and not upon intrinsic properties of the materials, which would have produced the opposite result. By comparative analysis of these dynamometer measurements and considerations of the lens design and elastic properties (including memory) of the component materials, we can calculate the stresses upon the zonular and capsular bag structures during and after IOL implantation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.