Abstract
The complement activation potential of surface modified (passivated) poly(methyl methacrylate) (PMMA) intraocular lenses (IOLs) with polypropylene loops was compared to that of standard PMMA IOLs with polypropylene loops. Both lens types were incubated in human sera for six hours and then the C3a and C5a levels were measured by radioimmunoassay. Sera incubated with either IOL type generated significantly higher levels of C3a and C5a than control sera incubated without any IOL. The amount of C3a and C5a generated by the passivated PMMA IOLs was comparable to the levels generated by the standard PMMA IOLs. The results of this study show that surface passivated PMMA IOLs with polypropylene loops activated complement to the same level as standard PMMA IOLs with polypropylene loops.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.