Abstract

In this work we have considered metrological problems and measurement of magnetic parameters and presented methods of measuring effective magnetic anisotropy field HAeff and ferromagnetic resonance bandwidth ∆H in magnetically uniaxial hexagonal ferrites in the electromagnetic microwave working frequency range. The methods allow measuring HAeff in the 10–23 and 28–40 kE ranges and ∆H in the 0.5–5.0 range. One method (suitable for wavelength measurements in free space in the 3-mm wavelength range) has been implemented for the 78.33–118.1 GHz range. The other method (based on the use of microstrip transmission lines) has been implemented for the 25–67 GHz range. The methods have been tested for polycrystalline specimens of hexagonal barium and strontium ferrites with nominal composition or complex substituted and having high magnetic texture. The measurement results have been compared with those obtained using conventional measurement methods and spherical specimens. Our methods prove to be highly accurate and reliable.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.