Abstract

The correlation of impedance cardiography against electromagnetic flow measurement was studied in the pig. The conditions tested were normal circulation, hypo- und hypervolemia and acute pulmonary hypertension as well as the influence of respiration and electrodes location. Correct quantification of cardiac output was obtained only in normovolemic circulation during exspiration (r = 0,96, m = 1,15, b = 0,04).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call