Abstract

A new protein assay system for the antigen-antibody interaction was developed by immobilization of carcinoembryonic antibody (anti-CEA) onto magnetic-core/gold-shell nanoparticles-functionalized biomimetic interface on multiporous polythionine modified magnetic carbon paste electrodes (MCPE). Differential pulse voltammetric (DPV) technique was employed to investigate the antigen-antibody interaction in pH 6.8 acetate acid buffer solution after incubation with various CEA samples for 50 min at room temperature. The peak currents decreased with increased CEA concentration, and were proportional to the CEA concentration in the range of 1.5-60 ng/ml with a detection limit of 0.3 ng/ml at a signal-to-noise ratio of 3. Moreover, the selectivity, reproducibility and stability of the proposed immunoassay system were acceptable. Compared with the conventional immunoassays, the developed immunoassay system was simple and rapid without multiple labeling and separation steps. Importantly, the proposed methodology would be valuable for diagnosis and monitoring of carcinoma and its metastasis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.