Abstract

Metal oxide-filled reactors constructed with ceramic tubes or fused silica capillary are widely used for combustion in gas chromatography combustion isotope ratio mass spectrometry (GCC-IRMS). However, they tend to be easily cracked or broken and prone to leaks at operating temperatures of ∼950 °C. Here we introduce a modified commercially available catalytic combustion/reduction methanizer to quantitatively convert organics to CO2 for δ13C analysis while retaining chromatographic resolution. These modified "ARC" reactors operate with a transition-metal catalyst that requires a flowing O2 gas to enable complete conversion to CO2 at lower temperature (620 °C) with acceptable reactor life, reduced complexity, and improved robustness. Performance of two versions of the ARC reactors with different combustion volumes was characterized by analysis of steroid and alkane isotopic standard materials. Linearity of steroid isotopic standards ranged from 0.02 to 0.60 ‰/V in the range of 25 to 200 ng of each steroid injected. Precisions and accuracies of measurements for steroids and alkanes had average standard deviations of SD(δ13C) less than ±0.18 ‰ and average accuracy of better than 0.19 ‰ δ13CVPDB. Peak width expansion within both devices were similar to that in traditionally used metal oxide reactors. These data demonstrate for the first time that novel combustion schemes enable operation at lower temperatures as an alternative approach comparable to high temperature techniques to yield high precision δ13C data with GCC-IRMS.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call