Abstract

This study was directed towards investigating suitable compounds to be used as stable isotope reference materials for gas chromatography combustion isotope ratio mass spectrometry (GC-C-IRMS) calibration. Several compounds were selected from those used in the 'Grob-test' mixture. Oxygen- and nitrogen-containing substances were added to these compounds to allow the mixture to be used as a possible multi-isotopic calibration tool for 2H/1H, 13C/12C, 15N/14N and 18O/16O ratio determinations. In this paper we present the results of delta13C measurements performed by the consortium of the five laboratories taking part in this inter-calibration exercise. All the compounds were individually assessed for homogeneity, short-term stability and long-term stability by means of EA-IRMS, as required by the bureau communitaire de reference (BCR) Guide for Production of Certified Reference Materials. The results were compared then with the GC-C-IRMS measurements using both polar and non-polar columns, and the final mixture of selected compounds underwent a further certification exercise assessing limits of accuracy and reproducibility under specified GC-C-IRMS conditions.