Abstract

Wireless telemetry is crucial for long-term implantable neural recording systems. RF-encoded neurological signals often require high data-rates to transmit information from multiple electrodes with a sufficient sampling frequency and resolution. In this work, we quantify the effects of interferers and tissue attenuation on a wireless link for optimal design of future systems. The wireless link consists of an external receiver capable of demodulating FSK/OOK transmission at speeds up to 8 Mbps, with <1e-5 bit-error rate (BER) without error correction, and a fully implanted transmitter consuming about 1.05 mW. The external receiver is tested with the transmitter in vivo to show demodulation efficacy of the transcutaneous link at high data-rates. Transmitter/Receiver link BER is quantified in typical and controlled RF environments for ex vivo and in vivo performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.