Abstract

A low density lipoprotein (LDL) cholesterol testing device, structured with serial reflected face-to-face mirror (SRM) allowing spectrophotometry measurements, is presented. The spectrophotometry has been employed to measure the amount of light that a sample absorbs, but it generally should have had path length longer than 10 mm to secure enough sensitivity. Such requirement of path length has often been problematic in implementing a thin type of lab on a disc (LOD). We developed the SRM system which was implemented in a detection chamber with 1.4mm thickness, providing path length longer than 10mm, and thus straightforwardly being applicable to LOD as thin as a compact disc. The experimental results show that the SRM system gives not only a much thinner design compared to the conventional spectrophotometry-based LOD but also a comparable performance to already commercialized spectrometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.