Abstract
Aim: A novel LC-MS/MS method using a surrogate matrix and derivatization with fluorescamine was developed and validated for simultaneous quantification of asymmetric dimethyl arginineand symmetric dimethyl arginine. Methods & results: Asymmetric dimethyl arginine, symmetric dimethyl arginineand corresponding internal standards were extracted using protein precipitation and derivatization with fluorescamine followed by SPE. Derivatives were analyzed by turbo ion spray LC-MS/MS in the positive ion mode. Methodology was successfully transferred across multiple preclinical species and utilized in the support of several investigative studies. Conclusion: A new LC-MS/MS analytical methodology that utilizes a surrogate matrix and derivatization with fluorescamine was successfully developed and validated.