Abstract
The use of digital subtraction techniques combined with fluoroscopy has rekindled interest in arteriography using intranvenous injections of contrast media. A new method is proposed for intravenous angiography in which an x-ray source and xenon detector array from a computed tomographic (CT) scanner are used to scan a region of interest to produce projection image. In order to provide adequate visualization of small concentrations of iodine in blood vessels, a subtraction scheme is used to remove the contribution from overlapping soft tissue and bone. Initial experiments with a temporal subtraction algorithm on phantoms have demonstrated the ability to image simulated blood vessels of 1.7-mm diameter containing dilute diatrizoate with an iodine concentration of 3.7 mg/cc, at an exposure of less than 100 mR. Vascular structures 5-8 mm in diameter have been imaged in dogs with iodine concentrations of less than 37 mg/cc using temporal subtraction. Principal advantages of the method over other film or fluoroscopic subtraction techniques are: 1) wide dynamic range an low noise of the (CT) detectors, providing excellent iodine sensitivity; 2) high scatter rejection; and 3) efficient utilization of x-ray dose.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.