Abstract
Interlaboratory comparisons are essential to bringing emerging technologies into biopharmaceutical industry practice and regulatory acceptance. As a result, an international team including 12 laboratories from 10 independent biopharmaceutical companies in the United States and Switzerland was formed to evaluate the precision and robustness of capillary isoelectric focusing (CIEF) to assess the charge heterogeneity of monoclonal antibodies. The different laboratories determined the apparent pI and the relative distribution of the charge isoforms of a representative monoclonal antibody (rMAb) sample using the same CIEF method. Statistical evaluation of the data was performed to determine within and between-laboratory consistencies and outlying information. The apparent pI data generated for each charge variant peak showed very good precision between laboratories with percentage of RSD values of ≤0.5%. Similarly, the percentage of RSD for the rMAb charge variants percent peak area values are ≤4.4% across different laboratories with different analysts using different lots of ampholytes and multiple instruments. Taken together, these results validate the appropriate use of CIEF in the biopharmaceutical industry in support of regulatory submissions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.