Abstract

Abstract Integrated PVDF pyroelectric radiation detectors have been made, using a simple technique to interface the pyroelectric material with the silicon substrate. The silicon substrates incorporate the electronics for a 8x1 array and a 2x2 array. The relevant sensor parameters voltage responsivity, detectivity and the thermal crosstalk are measured within a frequency range from 0.01 Hz to 5000 Hz. With the 400 μm thick silicon substrate the voltage responsivity is typically 100 V/W, the maximum detectivity is 7×106 cm□Hz/W at a frequency of 30 Hz. An improvement of both voltage responsivity and detectivity is possible by reducing the heat loss to the silicon substrate. This can be realized by a reduction of the substrate thickness, at least underneath the detector area. A further improvement is possible by a reduction of the capacity of the electronic circuit on the silicon chip and by the use of ferroelectric materials with higher pyroelectric coefficients, for example TGS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.