Abstract

C, N, O, F and P can be analyzed by instrumental photon activation analysis (IPAA) including decay curve analysis. The interference of 30P (T1/2 = 149.9 s) by 15O (T1/2 = 122.2 s) can be ruled out by direct positron measurement making use of the largely different maximum β+-energies of both nuclides (3.24 MeV and 1.73 MeV, respectively). Interference by carbon (11C) can be avoided by sub-threshold activation with 17 MeV bremsstrahlung. The short half-life of 30P allows a high productivity of the method. Reliability was demonstrated in the range of 0.2%–2% P (detection limit = 40 μg/g). Analysis of a certified reference material (BCR-CRM 063) yielded excellent agreement with the certified data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.