Abstract

In order to precisely measure the ion parameters in a microwave electron cyclotron resonance plasma using an ion sensitive probe, the dependences of the current-voltage (I-V) characteristics on the shielding height (h) and the potential difference between inner and outer electrodes (VB) have been investigated at different working pressures of 0.03 Pa and 0.8 Pa. Results show that the I-V curves at higher pressure are more sensitive to the variation of h than those at lower pressure. The influence of VB on ion temperature (Ti) measurement becomes more prominent when the pressure is increased from 0.03 Pa to 0.8 Pa. Under both pressures, the optimized h is obtained at the condition where the current reaches zero in the positive voltage region with a suitable VB of −1.5 V because of effective shielding of the electron E×B drift.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.