Cortical neural prostheses require chronically implanted small-area microelectrode arrays that simultaneously record and stimulate neural activity. It is necessary to develop new materials with low interface impedance and large charge transfer capacity for this application and we explore the use of conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT) for the same. We subjected PEDOT coated electrodes to voltage cycling between -0.6 and 0.8 V, 24 h continuous biphasic stimulation at 3 mC/cm² and accelerated aging for four weeks. Characterization was performed using cyclic voltammetry, electrochemical impedance spectroscopy, and voltage transient measurements. We found that PEDOT coated electrodes showed a charge injection limit 15 times higher than Platinum Iridium (PtIr) electrodes and electroplated Iridium Oxide (IrOx) electrodes when using constant current stimulation at zero voltage bias. In vivo chronic testing of microelectrode arrays implanted in rat cortex revealed that PEDOT coated electrodes show higher signal-to-noise recordings and superior charge injection compared to PtIr electrodes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Neural Systems and Rehabilitation Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.