Abstract
Nerve cuff electrodes for peripheral nerve prostheses are required chronically implanted electrodes which simultaneously stimulate and record nerve activity. It is inevitable challenge to investigate electrode material with low interfacial impedance and enhanced charge transfer capacity. In this study, stimulus nerve cuff electrodes on polyimide with Pt, conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT), black Pt, and IrOx were fabricated and characterized. The electrochemical properties were investigated using cyclic voltammetry, electrochemical impedance spectroscopy, and voltage transient measurements. From th experimental results, stimulus nerve cuff electrodes with black Pt showed the highest charge delivery capacity (80 times higher than Pt), charge injection capacity (6 times higher than Pt), and lowest interfacial impedance (3.8 times lower than Pt).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.