Abstract

Protein assays commonly used to evaluate reusable device cleanliness do not always accurately measure the low concentrations of protein that are expected on reusable medical devices after processing. Methods often are adapted to provide an estimation of protein concentration; however, sensitivity issues in the portion of standard curves at the acceptance criteria of 6.4 μg/cm2 protein have been reported. Using analytical validation criteria, method improvements for the micro-bicinchoninic acid assay for protein residuals are demonstrated by incorporating a standard addition method, increasing the well volume, and changing the working reagent ratio. These improvements increased method sensitivity and accuracy in the reliable detection of protein levels for device cleaning validations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.