Abstract
An improved circuit consisting of a simple voltage follower for the measurement of electron energy distribution function in a plasma has been proposed. This circuit makes it possible to eliminate the fluctuation of plasma space potential, the deviation of dc probe voltage caused by the probe current drawing across the plasma resistance, and the attenuation of the second derivative signal during the measurement, simultaneously. Both the probe circuit considerations and the experimental results are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.