Abstract

In these immunoturbidimetric assays developed for apolipoproteins A, AI, AII, and B we used the Cobas-Bio centrifugal analyzer and commercially available antisera and calibrators. Within-day and within-run precision ranged from 1.0 to 5.8% for the four assays. The addition of bilirubin, up to 150 mg/L, or hemoglobin, 50 g/L, to sera did not interfere with the measurement of these apolipoproteins. Comparison of results of the immunoturbidimetric assays with those of radial immunodiffusion assays yielded slopes of 1.002, 0.822, and 0.957 and intercepts of -0.02, 0.47, and 0.03 g/L for apolipoproteins B, A, and AI, respectively. Reference intervals for the four apolipoproteins were determined by using sera from 193 apparently healthy individuals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.