Abstract

We developed an immunoturbidimetric assay for prealbumin on the Cobas Bio centrifugal analyzer and compared results from this assay with those from a rate nephelometric assay (Beckman Instruments Inc.) and a radial immunodiffusion kit (Behring Diagnostics). All three assays were evaluated for precision, linearity, and correlation to each other for analysis of sera from pediatric patients. All assays gave similar results for patients' samples. Values were higher by the radial immunodiffusion assay than by the other two methods, which gave similar results for the same specimens. We conclude that the immunoturbidimetric and rate nephelometric assay for prealbumin are acceptable alternatives for quantifying prealbumin in serum and also have a faster turnaround time than radial immunodiffusion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.