Abstract

PurposeTo characterize the performance of the Precise Image (PI) deep learning reconstruction (DLR) algorithm for abdominal Computed Tomography (CT) imaging. MethodsCT images of the Catphan-600 phantom (equipped with an external annulus) were acquired using an abdominal protocol at four dose levels and reconstructed using FBP, iDose4 (levels 2,5) and PI (‘Soft Tissue’ definition, levels ‘Sharper’,‘Sharp’,‘Standard’,‘Smooth’,‘Smoother’). Image noise, image non-uniformity, noise power spectrum (NPS), target transfer function (TTF), detectability index (d’), CT numbers accuracy and image histograms were analyzed. ResultsThe behavior of the PI algorithm depended strongly on the selected level of reconstruction. The phantom analysis suggested that the PI image noise decreased linearly by varying the level of reconstruction from Sharper to Smoother, expressing a noise reduction up to 80% with respect to FBP. Additionally, the non-uniformity decreased, the histograms became narrower, and d’ values increased as PI reconstruction levels changed from Sharper to Smoother. PI had no significant impact on the average CT number of different contrast objects. The conventional FBP NPS was deeply altered only by Smooth and Smoother levels of reconstruction. Furthermore, spatial resolution was found to be dose- and contrast-dependent, but in each analyzed condition it was greater than or comparable to FBP and iDose4 TTFs. ConclusionsThe PI algorithm can reduce image noise with respect to FBP and iDose4; spatial resolution, CT numbers and image uniformity are generally preserved by the algorithm but changes in NPS for the Smooth and Smoother levels need to be considered in protocols implementation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call