Abstract

Abrasion-induced insulation breach is a common failure mode of silicone-body, transvenous, implantable cardioverter defibrillator leads. It is caused either by external compression or internal motion of conducting cables. The present method of monitoring lead integrity measures low frequency conductor impedance. It cannot detect insulation failures until both the silicone lead body and inner fluoropolymer insulation have been breached completely, exposing conductors directly to blood or tissue. Thus the first clinical presentation may be either failure to deliver a life-saving shock or painful, inappropriate shocks in normal rhythm. We present a new method for identifying lead failure based on high frequency impedance measurements. This method was evaluated in 3D electromagnetic simulation and bench testing to identify insulation defects in the St. Jude Medical Riata® lead, which is prone to insulation breach.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.