Abstract
An automatic analyser based on high-performance liquid chromatography has been developed for the quantification of biogenic guanidino compounds in human physiological fluids. Fourteen guanidino compounds are mutually separated within 35 min on a cation-exchange column with a stepwise gradient elution of pH and/or ionic strength in the mobile phase and then converted automatically to their fluorescent derivatives with benzoin. The method in this system is simple, rapid and sensitive; the lower limits of detection are 5–50 pmol for monosubstituted guanidino compounds, 1 nmol for creatine and 20 nmol for creatinine in 100 μl of injection volume.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.