Abstract

A fully automated pretreatment system based on sequential injection solid-phase extraction (SPE) coupled to a graphite furnace atomic absorption spectrometer (Auto-Pret-GFAAS system) was developed to determine trace amounts of cadmium and lead. A handmade minicolumn packed with a chelating resin was used for the preconcentration of both metals. All protocol for the on-line SPE method was controlled by home-made software. A trigger switch that was placed next to the graphite furnace was used to synchronize the home-made software with built-in software in the GFAAS. One milliliter of sample solution was flowed through into the minicolumn, the analytes were collected and concentrated on the solid phase, and the analytes were eluted with nitric acid (3 M), and detected by GFAAS. The limits of detection (3sigma) for cadmium and lead were 0.20 and 2.6 ng L(-1), respectively. The sample throughput was 47 h(-1) for 1 mL sample loading. The proposed sensitive method with the original Auto-Pret-GFAAS system was applied to the determination of cadmium and lead in tap water and in leached solutions from ceramic ware using 4% acetic acid.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.