Fine needle-type glucose sensors with an outer diameter of less that 0.2 mm were fabricated using a low-cost biodegradable γ-polyglutamic acid (PGA) not only as a glucose oxidase (GOx) immobilizing material, but also as a permselective inner membrane. PGA film was prepared on a cellulose acetate-coated Pt-Ir wire electrode, and GOx was covalently immobilized on its surface using water-soluble carbodiimide. The obtained electrode was practically not affected by the existence of electroactive compounds, and provided long-term stability for approximately one month. It also functioned in horse serum with a good linear relationship between the current and glucose concentrations within the range of 2.8 to 22.4 mM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.