Abstract
We present a CMOS-based high-density microelectrode array (HD-MEA) system that enables high-density mapping of brain slices in-vitro with multiple readout modalities. The 4.48×2.43 mm2 array consists of 59,760 micro-electrodes at 13.5 µm pitch (5487 electrodes/mm2). The overall system features 2048 action-potential, 32 local-field-potential and 32 current recording channels, 32 impedance-measurement and 28 neurotransmitter-detection channels and 16 voltage/current stimulation channels. The system enables real-time and label-free monitoring of position, size, morphology and electrical activity of brain slices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Solid-State Sensors, Actuators and Microsystems Conference : [proceedings]. International Conference on Solid-State Sensors, Actuators, and Microsystems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.