Abstract

An improved method for the determination of selected organochlorine pesticides (OCPs) and polychlorinated biphenyls (PCBs) in human serum was developed. The method requires low volume of serum (500 μl) and 48–96 samples per day can be prepared by one analyst without special automatic equipment. Initial extraction was performed using 96-well solid-phase extraction disk plates and was followed by a clean-up with silica gel/sulfuric acid. Different denaturation, elution and clean-up conditions were tested. Quantification was carried out by gas chromatography equipped with electron capture detector (GC-ECD) or mass spectrometer (GC–MS). Recoveries of PCB congeners 28, 52, 101, 118, 138, 153 and 180 and OCPs HCB, β-HCH, p, p′-DDE and p, p′-DDT at two spiking levels ( n = 8) varied from 57 to 120%, and intra-day relative standard deviation from 1 to 11%, both depending on spiking level and compound. Inter-day relative standard deviation was <15% in all cases. Limit of quantification (LOQ) for these PCBs ranged from 0.08 to 0.13 ng/ml and for these OCPs from 0.16 to 0.40 ng/ml. The optimized method was applied to the analysis of 1000 serum samples from different places of Spain.