Abstract
We describe specifications for high-purity 4-nitrophenol, which is suitable for spectrophotometric standardization. Such a reference material is needed in clinical enzymology to establish the proper molar absorptivity of 4-nitrophenol under final reaction conditions, particularly for measuring alkaline phosphatase activity in human serum. Some lots of 4-nitrophenol available commercially met these specifications, but several did not. The latter can be purified to meet our specifications by recrystallization or sublimation. The molar absorptivity of 4-nitrophenol (35 mumol/L) IN 10 mmol/L NaOH at 25 degrees C at 401 nm is 18380 +/- 90 L.mol-1.cm-1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.