Polypeptides subjected to high-performance liquid chromatography on a TSK G2000SW column in acidic mobile phases of low ionic strength underwent a combination of size exclusion and ion exclusion from the matrix. The distribution coefficient ( K d) of each polypeptide was dependent upon the concentration of ion-pairing agent present and increased as the concentration of ion-pairing agent increased. The fractionation range of the column could thus be manipulated by altering the concentration of acid in the mobile phase. The nature of the ion-pairing acid, in terms of hydrophilic or hydrophobic ion pairing, also had an effect on the chromatography, such that hydrophobic ion-pairing agents lead to increased K d values, especially in the case of the most basic polypeptides. To account for these results it is proposed that the TSK GSW matrix is cationic at low pH, and that ion pairing suppresses the ion exclusion experienced by cationic polypeptides under these conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.