Abstract

The efficacy of the GHOST proximity correction technique for achieving submicron linewidth control in electron beam lithography has been experimentally evaluated. Using the AT&T electron beam exposure systems, the GHOST parameters (correction dose, defocused beam diameter, and overlay accuracy) required for proximity exposure correction at 20 kV were determined via scanning electron microscopy and electrical linewidth measurement techniques. Optimum proximity correction conditions were established from a matrix of correction doses and correction beam diameters on two different substrates: chromium on glass and aluminum on silicon. On chromium, linewidth control of better than ±0.05 μm was obtained using a 3‐μm beam and a correction dose between 30% and 40% of incident, or with a 4‐μm beam and a 40% correction dose. Even better proximity correction was achieved on aluminum with a 5‐μm defocused beam and a 30% correction dose. In conjunction with the experimental data, a computer model was developed to inte...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.