Abstract

Diamond photoconductive detectors have been shown to detect fast neutrons with high gamma insensitivity. Depending on the application and the incident neutron energy, there are many possible choices when considering how diamond elements may be sized, arranged, and instrumented. As part of our design effort, we are using Geant4 and MCNP6.2 to simulate the effects of fast neutrons impinging on diamond detectors ranging in thickness from a few microns to a few hundred microns that are 4 mm on a side with intervening materials and other physical parameters. The models may be used to compare diamond detector measurements with incident neutrons ranging from ~1 to 14.1 MeV to better understand the nuclear and atomic physics effects contributing to an electronic signal. We are investigating pulse height, signal-to-noise ratio, and timing characteristics of prototype single-crystal chemical vapor deposition diamond detectors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.