Abstract

A single-crystal CVD (Chemical Vapor Deposition) diamond detector was used to measure γ rays in order to assess its performance in terms of energy resolution and linearity. For this purpose, 57Co, 133Ba, 22Na, 207Bi and 137Cs γ sources were used. Electrons scattered by the backward Compton process were detected in the diamond, in coincidence with (backscattered) γs measured in a NaI detector, placed at 180° from the CVD diamond detector with respect to the source. The resulting calibration shows a linear dependence of the charge deposited in the diamond and a resolution of about 24keV FWHM for the energy of the incident γs between 40keV (57Co) and 477keV (137Cs), comparable with the resolution of our electronic chain.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.