Abstract

A gas chromatographic system for the evaluation of linear temperature-programmed retention indices allowing n-alkanes to be adopted as the reference retention markers for any type of analyte, irrespective of the atoms present in their molecules, is described. It is based on the simultaneous use of two different detectors (a flame ionization detector and a specific detector suitable for the sample components), both connected (in parallel) to the same column outlet. The performance of this system has been tested by measuring the retention indices of fifteen chlorinated pesticides under conditions of linear programming temperature, by adopting an electron-capture detector as the specific detector. The reliability of the retention indices thus determined has been proven by verifying that they can be reproduced under different chromatographic conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.