Abstract

ObjectivesThe purpose of this study is to determine the feasibility of polyetheretherketone-based dental implants, and analyze the stress and strain around different kinds of dental implants by finite element analysis. MethodsThe radiographic data was disposed to models in Mimics 19.0. 3D models of implants, crowns and jawbones were established and combined in SolidWorks 2018. Appling axial and oblique loads of 100 N, cloud pictures were exported in Ansys Workbench 18.0 to calculate and analyze the stress and strain in and around different implants. ResultsOblique load tended to deliver more stress to bone tissue than axial load. The uniformity of stress distribution was the best for 30% short carbon fiber reinforced polyetheretherketone implants at axial and buccolingual directions. Stress shielding phenomenon occurred at the neck of 60% continuous carbon fiber reinforced polyetheretherketone and titanium implants. Stress concentration appeared in PEEK implants and the load of bone tissue would aggravate. Conclusions30% short carbon fiber reinforced polyetheretherketone implants demonstrate a more uniform stress distribution in bone-implant contact and surrounding bone than titanium. Stress shielding and stress concentration may be avoided in bone-implant interface and bone tissue. Bone disuse-atrophy may be inhibited in PEEK-based implants.