Abstract

A field-size effect of physical doses was studied in scanning irradiation with carbon ions. For the target volumes of 60 x 60 x 80, 40 x 40 x 80, and 20 x 20 x 80 mm3, the doses along the beam axis within the spread-out Bragg peaks reduced to 99.4%, 98.2%, and 96.0% of the dose for the target of 80 x 80 x 80 mm3, respectively. The present study revealed that the observed reductions can be compensated for by adopting the three-Gaussian form of lateral dose distributions for the pencil beam model used in the treatment planning system. The parameters describing the form were determined through the irradiation experiments making flat concentric squared frames with a scanned carbon beam. Since utilizing the three-Gaussian model in the dose optimization loop is at present time consuming, the correction for the field-size effect should be implemented as a "predicted-dose scaling factor." The validity of this correction method was confirmed through the irradiation of a target of 20 x 20 x 80 mm3.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.