Abstract

Introduction We present a mathematical method to adjust the leaf end position for dose calculation correction in the carbon ion radiation therapy treatment planning system. Methods and Materials A straggling range algorism of 400 MeV/n carbon ion beam in nine different multileaf collimator (MLC) materials was conducted to calculate the dose 50% point to derive the offset corrections in the carbon ion treatment planning system (ciPlan). The visualized light field edge position in the treatment planning system is denoted as Xtang.p, and MLC position (Xmlc.p) is defined as the source to leaf end midpoint projection on axis for monitor unit calculation. The virtual source position of energy at 400 MeV/n and straggling range in MLC at different field sizes were used to calculate the dose 50% position on axis. On-axis MLC offset (correction) could then be obtained from the position corresponding to 50% of the central axis dose minus the Xmlc.p. Results The exact MLC position in the carbon ion treatment planning system can be used as an offset to do the correction. The offset correction of pure tungsten is the smallest among the others due to its shortest straggling range of carbon ion beam in MLC. The positions of 50% dose of all MLC materials are always located in between Xtang.p and Xmlc.p under the largest field of 12 cm by 12 cm. Conclusions MLC offset should be adjusted carefully at different field sizes in the treatment planning systems especially of its small penumbra characteristic in the carbon ion beam. It is necessary to find out the dose 50% position for adjusting MLC leaf edge on-axis location in the treatment planning system to reduce dose calculation error.

Highlights

  • We present a mathematical method to adjust the leaf end position for dose calculation correction in the carbon ion radiation therapy treatment planning system

  • We illustrate the specific issues to carry out dose calculation of a rectangular end multileaf collimator (MLC) system with an offset correction in a carbon ion beam

  • This work presented here was performed with 400 MeV/n on a carbon ion therapy facility established by the Institute of Modern Physics (IMP), China

Read more

Summary

Introduction

We present a mathematical method to adjust the leaf end position for dose calculation correction in the carbon ion radiation therapy treatment planning system. A straggling range algorism of 400 MeV/n carbon ion beam in nine different multileaf collimator (MLC) materials was conducted to calculate the dose 50% point to derive the offset corrections in the carbon ion treatment planning system (ciPlan). The visualized light field edge position in the treatment planning system is denoted as Xtang:p, and MLC position (Xmlc:p) is defined as the source to leaf end midpoint projection on axis for monitor unit calculation. The characteristic of the projected light field edge locations and the definition of MLC position as BioMed Research International well as the dose 50% points in the photon treatment planning system need to be corrected before patients’ treatment monitor units are calculated. We illustrate the specific issues to carry out dose calculation of a rectangular end MLC system with an offset correction in a carbon ion beam

Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call