Abstract

In-vivo dosimetry is essential to verify the position and the intensity of the radiation therapy, such as cranio-spinal irradiation (CSI) and total body irradiation (TBI). Various kinds of devices, such as a thermo-luminesence dosimeters (TLDs), metal-oxide semiconductor field effect transistors (MOSFETs), semiconductor diodes, and gafchromic films, are used in in-vivo dosimetry, and these have their respective pros and cons. An optically-stimulated luminescent nanodot dosimeter (OSLnD) made of Al2O3: C was developed to measure the radiation dose during diagnostics, but it is now used for clinical purposes. In this study, the characteristics of the OSLnD, such as its dose rate dependency, dose linearity, angular dependency, and field junction, were investigated under a 6 MV X-ray beam. The OSLnD showed a linear response at doses from 20 to 300 cGy in the dose linearity test. Also, the dose rate dependency was shown to be less than 3%, angular dependency to be less than 2%. The experimental results proved the OSLnD to be useful for measurements of the external dose and for intensity modulated radiotherapy (IMRT) in clinical radiotherapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.