Abstract

The Multigap Resistive Plate Chambers (MRPCs) provide excellent timing as well as position resolutions at relatively low cost. Therefore, they can be used in medical imaging applications such as PET where precise timing is a crucial parameter of measurement. We have designed and fabricated several six-gap glass MRPCs and extensively studied their performance. In this paper, we describe the detector, the electronics and the data acquisition system of the setup. We present here the data analysis procedure and initial results of our studies to measure the absolute position of a radioactive source (22Na) using Time Of Flight (TOF) as well as the hit coordinate information and hence to demonstrate their potential applications in medical imaging. We also present the geant4 based simulation results on the efficiency of our detector as a function of the number of gaps and converter material thickness.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call