Abstract

The Multigap Resistive Plate Chambers (MRPCs) provide excellent timing as well as position resolutions at relatively lower cost. Therefore, they can be used in medical imaging applications such as PET where precise timing is a crucial parameter of measurement. We have designed and fabricated several six-gap glass MRPCs and extensively studied their performance. In this paper, we describe the detector, electronics and the data acquisition system of the setup. We present here the data analysis procedure and initial results of our studies to measure the absolute position of a radioactive source (\(^{22}Na\)) using time of flight (TOF) as well as X- and Y- hit position information and hence to demonstrate their potential applications in medical imaging.