Abstract

This paper presents the dosimetric quantities, realistic neutron calibration fields and spectrum-weighted response (SWR) of several neutron-measuring devices to provide guidance to health physicists and dosimetry specialists on the proper selection and use of survey meters in neutron fields of varying energy spectra. To yield the SWR, the detector response functions of 17 neutron-measuring devices were numerically folded with each of the produced calibration neutron spectra through the in-house developed software 'K-SWR'. The detectors' response functions were taken from the IAEA Technical Reports Series No. 403. Eight kinds of calibration neutron fields with high energy and high dose rates were produced using a proton accelerator (MC50 cyclotron) of Korea Institute of Radiological and Medical Sciences (KIRAMS). The fluence-average energy (Eave) varied from 3.8 to 16.9 MeV, and the ambient-dose-equivalent rate [H(*)(10) h(-1)] varied from 0.99 to 16.5 mSv h(-1).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.