Abstract

We built and tested a detector to measure the profile of fast-neutron beams delivered by the MC50 cyclotron at the Korea Institute of Radiological and Medical Science (KIRAMS). The core component of the detector is a 2 × 46 array of scintillation fibers. The light output of the scintillation fibers is transformed into a current signal by a 46-channel silicon photodiode and digitized by a current-mode signal processor. This scanning device was designed to cover a neutron beam area of 30 × 32 cm 2 . The detector was tested in a neutron beam delivered by the MC50 cyclotron at KIRAMS. We demonstrate that the detector can successfully measure the neutron beam profile at various beam currents from 10 to 20 μ A . The proposed neutron beam profile detector will be useful, for example, in radiotherapy applications with neutron intensities above 10 7 Hz / cm 2 .

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.