Abstract

The authors report the fabrication of long-range surface plasmon polariton biosensors consisting of thin narrow Au stripes embedded in a low refractive index fluoropolymer with etched fluidic channels. The fabrication process incorporates a sacrificial SiO2 channel etch stop layer to avoid waveguide damage and prevent channel over-etching, and an Al etch mask to minimize thermal cracking of the fluoropolymer during channel etching. Process details are reported along with fabrication results, and the optical and biosensing performance of the fabricated devices are demonstrated. The biosensors were tested for bulk sensitivity and bovine serum albumin (BSA) protein adsorption producing notable changes for solutions with different refractive indices (increment of 2 × 10−3) and a change in insertion loss of 0.1 dB as a monolayer of BSA forms on the device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.