Abstract
We report here the design, fabrication, and testing of an integrated thermocouple gauge. The gauge uses a thin-film Cr heater and a thin-film Cu–Cr differential thermocouple on a 5-×3-mm glass substrate. The thermocouple measures the difference between the heater and the substrate temperature. Calibration of the thin-film Cu–Cr thermocouple gauge was done by using a standard K-type thermocouple as reference. At a constant difference temperature the measured power input to the heater is a function of the surrounding vapor pressure. The thermocouple gauge was calibrated using a McLeod gauge with He, Ar, N2, H2O, and CO2 gases. Our measurements show that the range of the gauge can possibly be extended to the μTorr range. This gauge uses a standard thin-film processing technique for fabrication resulting in a lower cost of production. Also, small size makes this gauge versatile and unique in comparison to many commercial gauges.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.