Abstract
The performance of an epithermal neutron (0.5eV<En<10keV) flux monitor designed for boron neutron capture therapy (BNCT) was experimentally studied by using a prototype monitor in an appropriate neutron field at the intense deuterium-tritium neutron source facility OKTAVIAN of Osaka University, Japan. It was convinced from the experimental results that the developed monitor worked well and the epithermal neutron fluxes in BNCT neutron sources can be measured within 5% by the monitor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.